Event summary: failure data files confirmed system notice (#(b)(4)) was received indicating that the safety system detected fluid in the catheter and stopped the injection for the date of the event.Upon visual inspection of balloon catheter 2af284/06208-(b)(4), results showed the device was intact with no apparent issues.Smart chip verification indicated the catheter was used for nine applications.The catheter failed the performance test due to system notice (#(b)(4)) triggered immediately.Dissection revealed dried blood on the guide wire lumen and inside the inner balloon.The guide wire lumen was kinked and twisted 1.04 inches from the tip.Pressure test revealed a leak that came from a breach inside the guide wire lumen at the kink point in conclusion, the reported issue (system notice# (b)(4)) was confirmed through testing and through data analysis.The balloon catheter failed the returned product inspection due to a guide wire lumen kink and breach.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, a system notice was received indicating that the safety system detected fluid in the catheter and stopped the injection.The balloon catheter and coaxial umbilical cable were both replaced and the case was completed with cryo.No patient complications have been reported as a result of this event.On 2018-03-07: the catheter was returned to the manufacturer, analyzed, and tested out of specification.
|