• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC CARTO 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC CARTO 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Catalog Number FG540000
Device Problem Improper Device Output (2953)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/09/2018
Event Type  malfunction  
Manufacturer Narrative
Product investigation summary: a biosense webster inc.(bwi) field service engineer (fse) visited the site to evaluate the unit.The fse performed cube and acl testing.The system passed the tests and could not duplicate issue.Fse provided case support and the system functioned normally.Issue was not duplicated.System ready for use.The device history record (dhr) review was performed by the manufacturer and no anomalies, which are related to the reported issue, were noted in manufacturing or servicing of this equipment.(b)(4).
 
Event Description
It was reported that a patient underwent a procedure with a carto 3 system wherein a mapshift issue occurred.The customer described the event as, the map geometry shifted and then eventually went back to normal.No adverse patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO 3 SYSTEM
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
BIOSENSE WEBSTER INC
33 technology drive
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER (ISRAEL) LTD.
4 hatnufah street
yokneam 20667 17
IS   2066717
Manufacturer Contact
gabriel alfageme
33 technology drive
irvine, CA 92618
949789-868
MDR Report Key7394923
MDR Text Key104540961
Report Number2029046-2018-01400
Device Sequence Number1
Product Code DQK
UDI-Device Identifier10846835000870
UDI-Public10846835000870
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133916
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFG540000
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/09/2018
Initial Date FDA Received04/03/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-