• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL TRUESPAN 12 DEGREE PEEK; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL TRUESPAN 12 DEGREE PEEK; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 228151
Device Problems Misfire (2532); Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/06/2018
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
It was reported that the truespan 12 degree implant misfired on second deployment.Half the implant remained in the gun.Implant then pulled out of the gun when the surgeon went to remove the gun from the joint.Implant had to be cut out.No patient consequences.Action taken for procedure:implant remaining was cut out and removed from the joint.New truespan was opened and used.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate evaluation statement: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.Further, a review into the depuy synthes mitek complaints system revealed no similar/dissimilar complaint for this lot of devices with the product code that were released to distribution.A non-conformance search was performed for this part (228151) and lot (l642084) combination.No non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN 12 DEGREE PEEK
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle CH-24 00
SZ  CH-2400
MDR Report Key7395142
MDR Text Key104679271
Report Number1221934-2018-50401
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026012
UDI-Public10886705026012
Combination Product (y/n)N
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 03/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Catalogue Number228151
Device Lot NumberL642084
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/09/2018
Initial Date FDA Received04/04/2018
Supplement Dates Manufacturer Received10/16/2018
Supplement Dates FDA Received10/25/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age22 YR
-
-