Product event summary: the data files and balloon catheter, 2af284 with lot number 35135, were returned and analyzed.The data files showed at least thirteen applications were performed with this catheter.Also, the bin files confirmed system notice 50032 ¿the safety system has detected a compromised outer vacuum.¿ visual inspection of the balloon catheter showed the catheter was intact with no apparent issue.The performance test was not performed due to persistent system notice (b)(4).Dissection/ pressure testing revealed a leak in the inner balloon at the distal attachment with the catheter tip.Also, dissection showed the guide wire lumen kinked at 1.32 inches proximal from the tip.In conclusion, the balloon catheter failed the returned product inspection due to the inner balloon leak and the guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, a system notice was received indicating that the safety system detected a compromised outer vacuum.The balloon catheter was replaced with resolve, and the case was completed with cryo.No patient complications have been reported as a result of this event.The balloon catheter subsequently tested out of specification per the manufacturer's investigation.
|