No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative was unable to replicate the reported issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced.No parts have been received by the manufacturer for evaluation.
|
A site representative reported that while outside of procedure, when the navigation system was plugged into an outlet after moving it out of operating room (or), the system rebooted unexpectedly.Site was verifying instruments, the cable was unplugged from the outlet and replugged.When re-plugged, the system rebooted again.There was no patient present at the time of the issue.
|