• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US RIGIDFIX FEM 3.3MM S/T XPIN; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US RIGIDFIX FEM 3.3MM S/T XPIN; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 210133
Device Problems Migration or Expulsion of Device (1395); Separation Failure (2547)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
The affiliate reported via email during a knee procedure, the trocar stayed blocked within the sleeve after drilling.No patient consequence reported.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the complaint device was received and evaluated.The reported failure is confirmed, the sleeve and trocar are cold welded together.There are visible striation marks on the trocar.The location of the trocar interlocking pins is superior to the mating portion of the sleeve.If the pins are not seated within these groves then the trocar is spinning within the guide sleeve causing enough heat and friction to cause the welding of the two parts.Therefore, this type of failure has been attributed to user technique issue.The dhr review indicated that this batch of devices were processed without incident, therefore, there is no evidence of manufacturing anomalies on the records reviewed.A review into the mitek complaints system revealed no other complaints of any kind for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIGIDFIX FEM 3.3MM S/T XPIN
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
MEDOS SÀRL
rue du puits-godet 20
neuchâtel CH-20 00
SZ   CH-2000
Manufacturer Contact
jennifer lawrence
325 paramount drive
raynham, MA 02767
5089776860
MDR Report Key7399558
MDR Text Key104734812
Report Number1221934-2018-50302
Device Sequence Number1
Product Code HTY
UDI-Device Identifier10886705000890
UDI-Public10886705000890
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K974341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Catalogue Number210133
Device Lot NumberL175098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/11/2017
Initial Date FDA Received04/05/2018
Supplement Dates Manufacturer Received04/09/2018
Supplement Dates FDA Received04/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-