• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/14/2018
Event Type  malfunction  
Manufacturer Narrative
Unique device identifier (udi) is unavailable.Device manufacture date is unavailable.No parts have been received by the manufacturer for evaluation.Part not returned for analysis.
 
Event Description
A manufacturer representative reported that, while in a functional endoscopic sinus surgery (fess), the navigation system became unresponsive without prompt from the user.The representative reported that the navigation system was restarted without resolution as the navigation system became unresponsive at the navigation task in the application software.It was noted that a third restart restored functionality.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no reported impact on patient outcome.No additional information was provided.
 
Manufacturer Narrative
Additional information: device manufacture date and unique device identification (udi) provided.Correction: product and related fields updated to proper value.The computer for the navigation system was returned to the manufacturer for analysis.The computer was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Manufacturer Narrative
Correction: a medtronic representative went to the site to test the equipment.Testing revealed that the computer of the navigation system was replaced to restore functionality.The system then passed the system checkout and was found to be fully functional.The computer of the navigation system was returned to the manufacturer for evaluation.Testing found that the cpu or mb of the computer had a failure.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7400278
MDR Text Key104551798
Report Number1723170-2018-01463
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 06/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/14/2018
Initial Date FDA Received04/05/2018
Supplement Dates Manufacturer Received04/12/2018
05/08/2018
Supplement Dates FDA Received05/02/2018
06/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age83 YR
Patient Weight75
-
-