MEDTRONIC MEXICO ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number 990063-020 |
Device Problem
Knotted (1340)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/09/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product event summary: the device and data files were returned and analyzed.Data files showed that 5 injections were performed with a balloon catheter on the date of the event without any system notice triggered.Visual inspection of the mapping catheter showed a kink on the loop and that the pebax tube was ribbed.The catheter failed the return test due to a kink on the loop, damaged pebax and a loop knot.The product analysis findings did not indicate a manufacturing related defect.In conclusion, the reported mapping issue was confirmed through testing and not data analysis.The catheter failed the return product inspection due to a kink on the loop, ribbed pebax and a loop knot.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, the mapping catheter was ¿jammed¿ in the sheath before use.It was later noted that the mapping catheter end was getting stuck in the introducer when trying to place the y-connector.The mapping catheter was replaced with resolve.The mapping catheter was returned to the manufacturer, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
|
|