MEDTRONIC MEXICO ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 990063-020
Device Problem Knotted (1340)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/09/2017
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the device and data files were returned and analyzed.Data files showed that 5 injections were performed with a balloon catheter on the date of the event without any system notice triggered.Visual inspection of the mapping catheter showed a kink on the loop and that the pebax tube was ribbed.The catheter failed the return test due to a kink on the loop, damaged pebax and a loop knot.The product analysis findings did not indicate a manufacturing related defect.In conclusion, the reported mapping issue was confirmed through testing and not data analysis.The catheter failed the return product inspection due to a kink on the loop, ribbed pebax and a loop knot.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the mapping catheter was ¿jammed¿ in the sheath before use.It was later noted that the mapping catheter end was getting stuck in the introducer when trying to place the y-connector.The mapping catheter was replaced with resolve.The mapping catheter was returned to the manufacturer, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7401828
MDR Text Key104809359
Report Number9612164-2018-00648
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2018
Device Model Number990063-020
Device Catalogue Number990063-020
Device Lot Number212612510
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/06/2017
Initial Date FDA Received04/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age39 YR
-
-