Model Number MMT-723RNAS
Device Problems Crack (1135); Device Dislodged or Dislocated (2923)
Patient Problem Hyperglycemia (1905)
Event Date 03/25/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer¿s mother reported via phone call that they experienced high blood glucose and the insulin pump was damaged.The customer¿s blood glucose level was 426 mg/dl.The customer states the top cap part cracked and came off near the battery compartment.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device had broken battery tube threads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7401839
MDR Text Key104621418
Report Number3004209178-2018-62441
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169513143
UDI-Public(01)00643169513143
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723RNAS
Device Catalogue NumberMMT-723RNAS
Device Lot NumberA1723RNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2018
Initial Date Manufacturer Received 03/25/2018
Initial Date FDA Received04/05/2018
Supplement Dates Manufacturer Received07/31/2018
Supplement Dates FDA Received08/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age13 YR
Patient Weight110
-
-