Model Number 10280959 |
Device Problem
Insufficient Information (3190)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/21/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
|
|
Event Description
|
It was reported to siemens that a malfunction occurred while operating the artis zeego system.During an interventional procedure the store monitor was not functional.The patient was safely removed from the system and transferred to an alternate system where the procedure was completed.We are unaware of any impact to the state of health of the patient involved.
|
|
Manufacturer Narrative
|
Siemens has completed an investigation of the reported event.The root cause could not be determined as the replaced ias pc was not returned for investigation.The investigation of the log files showed that the ias pc could not connect to the ivs pc through the high speed link network (netlink disconnect) resulting in a high speed link error displayed on the control monitor and the examination room monitor.According to the log files the user continued to use the system performing fluoroscopic imaging.But since the network connection between the ias pc and the ivs pc failed the "store fluoro" option was not available on the graphical user interface in the control room.Therefore the user could not store the scene.The replaced ias pc was not returned for investigation.The most probable cause for a high speed link error is a failure of the network port on ias pc.This was confirmed during a simulation of the reported issue.On the affected system the ias pc was replaced.No systematic issue in the field was identified.No corrective action in the installed base is planned based on the present event.
|
|
Search Alerts/Recalls
|