MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problems Device Difficult to Program or Calibrate (1496); Operating System Becomes Nonfunctional (2996)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/14/2018
Event Type  malfunction  
Manufacturer Narrative
Analysis summary: analysis was unable to confirm the customer complaint of the unit rebooting itself, it was left on overnight with no faults observed.The link electronic module and the hard disk drive connectors were reseated and the power supply as preventive measures.It was noted that the hard disk drive health was at 100% and no errors were found in the parsing sheet.It was additionally noted that the system fan was noisy and it was also replaced.The circuit boards and mechanical parts were inspected, the hard drive reconfigured, the software reloaded and updated and the device then passed all final functional and systems tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer rebooted during a programming session.The programmer has been returned for service.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7412073
MDR Text Key105288835
Report Number2182208-2018-00384
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/14/2018
Initial Date FDA Received04/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-