SMITH & NEPHEW MEDICAL LTD. PICO WITH SOFT PORT 15X20CM; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66801362
Device Problem Pumping Problem (3016)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/01/2018
Event Type  malfunction  
Event Description
It was reported that the machine stopped after two days of treatment.No patient harm reported.
 
Manufacturer Narrative
Device was not evaluated due to it was not returned for investigation.Investigation results: we have now completed our investigation into the reported complaint.As of today, no product return or further information for this complaint has become available preventing a more thorough investigation.If the product becomes available in the future, this case will be reopened.A device history record review was carried out for the batch number supplied.This confirmed that the product had been manufactured in accordance with defined procedures and specifications and there was no record of any problems or deviations occurring during manufacturing of this product in relation to the reported issue.Each pump is functionally tested by the supplier before shipment to smith & nephew.Any pump which fails to meet the predetermined acceptance criteria is rejected.As this pump has been released for sale, it can be confirmed that the device has passed all functional testing.On this occasion we are unfortunately unable to reach a definitive root cause of the problem due to insufficient information available for the device in question, however smith and nephew are continually investigating ways to develop and improve our products and we will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PICO WITH SOFT PORT 15X20CM
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 8N
UK  HU3 28N
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 8N
UK   HU3 28N
Manufacturer Contact
sarah freestone
101 hessle road
hull HU3 2-8N
UK   HU3 28N
MDR Report Key7412827
MDR Text Key104957988
Report Number8043484-2018-00080
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K112127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66801362
Device Lot Number1723
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/01/2018
Initial Date FDA Received04/10/2018
Supplement Dates Manufacturer Received03/01/2018
Supplement Dates FDA Received06/09/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-