• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SINGAPORE OPERATIONS SENSIA DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SINGAPORE OPERATIONS SENSIA DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number SEDRL1
Device Problems Invalid Sensing (2293); Capturing Problem (2891)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/20/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during use of the implantable pulse generator (ipg) an inquiry was received regarding inconsistent and inaccurate capture management measurements compared to in clinic manual thresholds.Incoming information indicated that the right ventricular (rv) lead impedance was stable, and in the clinic, manual testing of the rv thresholds and sensitivity were stable as well.The caller was informed regarding the potential of the ipg not detecting the evoked sense response appropriately.Troubleshooting of the ipg was scheduled, and the ipg remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSIA DR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SG  486056
Manufacturer (Section G)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SG   486056
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7415461
MDR Text Key105242698
Report Number3008973940-2018-00499
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/28/2018
Device Model NumberSEDRL1
Device Catalogue NumberSEDRL1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/20/2018
Initial Date FDA Received04/10/2018
Date Device Manufactured07/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
-
-