Siemens evaluated the submitted reagent which showed no obvious signs of damage or degradation; individual cassette foil packages were properly sealed, containing the expected desiccant & micropipette.No patient sample or instrument was provided.Performance of the submitted cassettes was evaluated on two clinitek status+ instruments using a urine solution contrived to a 25 miu/ml hcg target, which is the concentration listed in the customer insert as being minimum for a positive determination, and a pooled negative urine.Six replicate test values were collected on each urine.Based on the data collected, the customer report of false negative hcg performance from clinitest hcg when tested on a clinitek status+ instrument was not observed.The root cause for the discordant results is unknown.
|