MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); No Display/Image (1183); Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 03/22/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A manufacturer representative went to the site to test the equipment.Testing revealed that the computer for the navigation system was replaced to restore functionality.The system then passed the system checkout and was found to be fully functional.The computer for the navigation system was returned to the manufacturer for evaluation.Testing found that the hard drive failed testing.
 
Event Description
A manufacturer representative reported that, while outside of a procedure, the navigation system displayed that there was no input detected after pressing delete to remove patient data from the computer.It was reported that the navigation system was restarted to restore functionality, but the navigation system still displayed that no input was detected.There was no patient present when this issue was identified.No additional information was provided.
 
Manufacturer Narrative
Device code was corrected.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7422971
MDR Text Key105207538
Report Number1723170-2018-01593
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 05/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/22/2018
Initial Date FDA Received04/11/2018
Supplement Dates Manufacturer Received03/22/2018
Supplement Dates FDA Received05/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-