The device was not returned to the manufacturer for physical evaluation, therefore the failure mode cannot be confirmed.However, an investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the device history record review confirmed the labeling, material, and process controls were within specification.
|
The device was returned to the manufacturer for physical evaluation.The failure mode was confirmed.An investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances noted during the manufacturing process.There were 9 halsted mosquito forceps returned in unused condition.The forceps were tested with artificial testing material for clamping; the forceps did not function as designed.When attempting to clamp the forceps, instead of gripping, they cut through the material.The complaint report has been confirmed; the root cause has been identified as a workmanship or material deficiency.Appropriate action has been implemented to rectify this manufacturing deficiency.
|