• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 03/21/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.Device manufacturing date is unavailable.A medtronic representative reported that the reported issue could be replicated when metal was brought near the system.The logs for the navigation system were reviewed by medtronic personnel.However, the logs provided no additional insight into the probable cause of the anomaly.
 
Event Description
A site representative reported that during functional endoscopic sinus surgery (fess), the navigation system became unresponsive.Site used another medtronic navigation system to complete the proceed.The procedure was completed with the use of navigation.No information was provided on delay to procedure.No known impact on patient outcome.
 
Manufacturer Narrative
Additional information: device manufacture date and unique device identification (udi) provided.Correction: product and related fields updated to proper value.The computer for the navigation system was returned to the manufacturer for analysis.The computer was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Manufacturer Narrative
Correction: a medtronic representative went to the site to test the equipment.Testing revealed that the computer of the navigation system was replaced to restore functionality.The system then passed the system checkout and was found to be fully functional.The analysis of the computer of the navigation system was completed by medtronic personnel.Testing found that the cpu or mb of the computer had a failure.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
jonathan resch
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7425360
MDR Text Key105439711
Report Number1723170-2018-01607
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/21/2018
Initial Date FDA Received04/12/2018
Supplement Dates Manufacturer Received04/12/2018
05/08/2018
07/26/2018
Supplement Dates FDA Received05/07/2018
06/04/2018
07/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-