Staff reported that after 7 hours on pump, an immediate high arterial pressure was measured, followed immediately by an audio alarm and pumping stopped on the rpm.Pump's power was recycled with same result.Alarm still present.Staff could not verify actual alarm.Staff clamped lines and started manual operation with emergency drive.Staff reports that turning emergency drive was extremely difficult.Excessive force had to be used to turn pump.Staff immediately reduced occlusion which resolved issue.Staff continued use with emergency drive until replacement pump was installed.No patient harm was reported.(b)(4).
|
Maquet medical systems, usa(importer)submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary ag kehler strasse 31, 76437 rastatt, germany.A follow-up medwatch will be submitted when additional information becomes available.Reference exemption # e2018002.(b)(4).Beside statement by the field service technician the belts have been returned to maquet laboratory (lce - life cycle engineering).According to investigation report, signed by lce on (b)(6) 2019, damaged belts (2 pc.) could be confirmed.One belt is heavily damaged.The trapezoid profile is broken in several places and the reinforcement fibers are partly torn.The probable root cause is a twisting of the belt due to a pre-existing defect, for example kinking during transport or deviations in the belt profile.The damage to the second belt is probably a consequential damage due to friction on the other damaged belt.Thus the failure could be confirmed.Most probable root cause could be determined as twisting of the belt due to a pre-existing defect, for example kinking during transport or deviations in the belt profile.
|