• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Patient Involvement (2645)
Event Date 03/23/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Device manufacturing date is unavailable.The software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.
 
Event Description
Medtronic representative reported that while outside of procedure, the navigation system unexpectedly exited when switching digital and analog c-arms in a 2d fluoro or computed tomography (ct) and fluoro procedure.The reported issue occurs when using a digital c-arm, proceed to navigate task and do not log out of the app or power down the system.Adding an analog c-arm to spine procedure, when acquiring empty and anatomy image.Selecting the activate button after this the application exits.It was reported that the issue also occurs when using an analog c-arm and switch to a digital c-arm.There was no patient present at the time of the issue.
 
Manufacturer Narrative
Correction: updated device information received including catalog number and udi number.Additional information: device manufacture date received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stephanie riley
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7428896
MDR Text Key105443286
Report Number1723170-2018-01623
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735665
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/23/2018
Initial Date FDA Received04/13/2018
Supplement Dates Manufacturer Received04/17/2018
Supplement Dates FDA Received05/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/21/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-