Catalog Number 04877802190 |
Device Problem
Low Test Results (2458)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred in (b)(6).Unique identifier (udi)#: (b)(4).
|
|
Event Description
|
The customer complained of discrepant results for 1 patient sample tested for roche cardiac d-dimer between two h232 instruments.The initial d-dimer result from a venous sample on h232 instrument serial number (b)(4) was 0.6 (unit of measure not provided)."almost simultaneously," the sample was repeated on a second h232 instrument and the result was 1.5.The internal qc failed on h232 instrument with serial number (b)(4) so the customer thinks the result of 0.6 is incorrect.The internal qc for the second h232 instrument passed.The patient was not adversely affected.The cobas h232 instrument nor a similar device is approved for distribution in the united states.The instrument and test strips were requested for investigation.
|
|
Manufacturer Narrative
|
The customer did not return the meter or the test strips.As no product was returned, the investigation could not be completed.The investigation was unable to find a definitive root cause.
|
|
Manufacturer Narrative
|
Relevant retention material roche cardiac d-dimer of lot 25712710 was measured on a cobas h232 at the investigation site with: two native blood samples and two spiked blood samples (c=0.60 g/ml and c=1.5 g/ml), each blood sample n=three test strips.Mean of the measurements on the cobas h232 at the investigation site: first native blood sample: 0.31 g/ml.Second native blood sample: 0.16 g/ml.First spiked blood sample (c=0.60 g/ml): 0.55 g/ml.Second spiked blood sample (c=1.5 g/ml): 1.30 g/ml.The results of all the measurements meet the requirements.
|
|
Search Alerts/Recalls
|