• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US OMNISPAN MENISCAL REPAIR 12DEG; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US OMNISPAN MENISCAL REPAIR 12DEG; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 228141
Device Problems Detachment Of Device Component (1104); Operating System Becomes Nonfunctional (2996)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/16/2018
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4) ¿ incomplete.The expiration date is not currently available.
 
Event Description
Casing damaged.The white casing is broken which lead to two baffles to fall off automatically and can not be used any more during in the meniscus prosthesis surgery.Changed another one to complete.Enclosed please find the defect photos.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was received and inspected.The complaint can be confirmed.It was observed that the casing which holds the implants on the needle was broken, and the implants were still attached to the needle via the orthocord.One of the implants appeared to be previously deployed.When the needle is inserted too forcefully into the tissue or at a depth that is too deep, the casing that holds the implants may be damaged therefore is a potential root cause for the reported failure.However, given the information provided we cannot discern a definite root cause for the reported failure.A device history record (dhr) review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.At this point, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNISPAN MENISCAL REPAIR 12DEG
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7434609
MDR Text Key106034536
Report Number1221934-2018-50472
Device Sequence Number1
Product Code MAI
UDI-Device Identifier10886705010035
UDI-Public10886705010035
Combination Product (y/n)N
PMA/PMN Number
K150209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2020
Device Catalogue Number228141
Device Lot NumberL374377
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2018
Initial Date Manufacturer Received 03/19/2018
Initial Date FDA Received04/16/2018
Supplement Dates Manufacturer Received05/31/2018
Supplement Dates FDA Received10/28/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age47 YR
Patient Weight92
-
-