• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND SYSTEM W/PINPOINT GT; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND SYSTEM W/PINPOINT GT; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770501
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will be evaluated.Results are expected soon.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
Per tm picture quality seems to have a film on it and had deteriorated.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: during evaluation, the reported issue with poor image quality could not be reproduced.The scanner was evaluated with the model 550 phantom (t11900271) and the image was found to be normal for a sr 8 scanner with the 32 mm probe.The review of the dhr, sub-assembly dhr, sub-assembly manufacturing, component records, manufacturing process changes, mrr/mrb, and internal rejects showed that the reported issue is not likely related to a manufacturing issue.No other issues with the equipment were found.The sr8 was serviced, tested, and returned to the customer.A history review of serial number dybyac505 showed one other similar complaint(s) from this serial number.The complaints for this serial number (dybyac505) have been reported from one u.S.Facility.
 
Event Description
Per tm picture quality seems to have a film on it and had deteriorated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND SYSTEM W/PINPOINT GT
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
DYMAX CORP.
141 zehner school road
zelienople PA 16063
Manufacturer Contact
sydney freckleton
605 n. 5600 w.
salt lake city, UT 84116
8015225992
MDR Report Key7435124
MDR Text Key105774319
Report Number3006260740-2018-00719
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00801741098338
UDI-Public(01)00801741098338
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770501
Device Catalogue Number9770501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2018
Is the Reporter a Health Professional? Yes
Event Location Hospital
Initial Date Manufacturer Received 03/30/2018
Initial Date FDA Received04/16/2018
Supplement Dates Manufacturer Received08/24/2018
Supplement Dates FDA Received08/28/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-