Model Number FUSION
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 03/27/2018
Event Type  malfunction  
Manufacturer Narrative
Device manufacture date is unavailable.A medtronic representative went to the site to test the equipment.The representative was unable to replicate the reported issue.It was noted that the representative suspected that the instrument was introduced too close to the patient tracker and the navigation system reverted.No additional information was provided.
 
Event Description
Medtronic received information that, while outside of a procedure, the application software would revert to the registration task when an instrument was brought within close proximity of the emitter.There was no patient present when this issue was identified.No additional information was provided.
 
Manufacturer Narrative
Additional information: unique device identification (udi) and device manufacture date updated to proper value.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
heather davis
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7437217
MDR Text Key105714431
Report Number1723170-2018-01656
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/27/2018
Initial Date FDA Received04/17/2018
Supplement Dates Manufacturer Received04/18/2018
Supplement Dates FDA Received05/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/21/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-