TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 816300
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/27/2018
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) verified the reported issue.He installed a loaner ccm.The unit operated to the manufacturer's specifications.The suspect unit was returned to the manufacturer for further evaluation.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit displayed a "disk boot failure" error message on the central control monitor (ccm).As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed the central control monitor (ccm) to power on to a black screen displaying 'disk boot failure, insert system disk and press enter'.The ccm battery measured 0 volts (v) (nominal voltages is three volts).After replacing the battery with a lab use only battery he observed the ccm to boot normally.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7440284
MDR Text Key106144729
Report Number1828100-2018-00193
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816300
Device Catalogue Number816300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/27/2018
Initial Date FDA Received04/18/2018
Supplement Dates Manufacturer Received06/11/2018
Supplement Dates FDA Received06/14/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-