The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital disposed of the device.
|
During preparation for a thrombectomy procedure on the back table, the indigo system separator d (sepd) was accidentally knocked off the table onto the floor.The sepd fell on the floor prior to use and therefore, it was not used in the procedure.The procedure was completed using an indigo system catd aspiration catheter (catd).
|