Model Number FUSION COMPACT |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/28/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device manufacturing date is unavailable.A medtronic representative went to site to test the equipment.Representative was unable to replicate the reported issue and the system was not showing any of the reported issue.The issue was has not occurred again.Representative reinstalled software as a precaution.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced.No parts have been received by the manufacturer for evaluation.
|
|
Event Description
|
Medtronic received information that during functional endoscopic sinus surgery (fess), the screen of the navigation system turned black when navigating.The system was rebooted but after rebooting the screen was flashing through configuration settings, starting and stopping and resetting.Rebooting the system with and without axiem box plugged did not resolve the issue.The procedure was completed without the navigation.There was no delay to procedure.No known impact on patient outcome.
|
|
Manufacturer Narrative
|
Additional information: device manufacture date provided.Correction: unique device identification (udi) updated to proper value.
|
|