Catalog Number 9735665
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/29/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The imaging system then passed the system checkout and was found to be fully functional.(b)(4).The software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.(b)(4).No parts have been received by the manufacturer for evaluation.(b)(4).
 
Event Description
Information was received from a site.They reported that while in a cranial resection procedure, the navigation system screen went black and displayed an error.The issue occurred when the site loaded the patient, moved to registration and started initializing.They restarted the system and went back to the registration screen.When they started initializing, it displayed the error again.The surgeon continued the case without the navigation system.There was no reported impact on patient outcome.There was a reported delay to the procedure of less than 1 hour due to this issue.No further information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
sarah olhoft
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7445004
MDR Text Key106033025
Report Number1723170-2018-01692
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735665
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/29/2018
Initial Date FDA Received04/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age64 YR
Patient Weight94
-
-