MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Unable to Obtain Readings (1516); Impedance Problem (2950)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/29/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received by a manufacture representative (rep) regarding an implantable neurostimulator (ins) for urinary dysfunction/sacral nerve stimulation and gastrointestinal/pelvic floor.Caller is reporting the following about impedances: intra-op abnormal impedances.They were with the patient and/or healthcare provider (hcp) to do further troubleshooting.The intra-op caller was getting 2767 repeating and ??? on case pairs while testing at 1.5v on 360 pulse width (pw).The call agency requested the rep to do motor testing; they tested pairs 0&3 and 1&2, but resulted in motor responses >1.5v (1.2v and >1.0v).The call agency then recommended closing and discussed environmental noise as possible reason for abnormal impedances.Troubleshooting resolved reported issue.Additional information received from a manufacturer representative (rep) who confirmed the information with the hcp; the cause of the impedance issue was not determined.No additional actions/interventions were taken to resolve the impedance issue.Before closing the patient, the rep said the impedance values were: 1378, ?, ?, ?, 2767, 2767, 2767, 2767, 2767, and 2767.No patient symptoms were reported and no further complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7450259
MDR Text Key106962587
Report Number3004209178-2018-08559
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00613994913654
UDI-Public00613994913654
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2019
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/29/2018
Initial Date FDA Received04/23/2018
Date Device Manufactured02/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age44 YR
-
-