• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37603
Device Problems High impedance (1291); Migration or Expulsion of Device (1395); Malposition of Device (2616)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2018
Event Type  Injury  
Manufacturer Narrative
Other relevant device(s) are: product id: neu_unknown_lead, serial/lot #: unknown.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturing representative about a patient with an implantable neurostimulator (ins) for dystonia.It was reported that the patient had high impedance on the right lead.The patient's lead was explanted and replaced.An x-ray of the system showed the connector was placed in the mastoid region, almost in the patient's neck.After lead replacement the impedance was fine.The patient is receiving effective therapy.It was unknown if there were any external or environmental factors that led to the event.No further complications were reported or anticipated.
 
Manufacturer Narrative
Implanted: explanted: (b)(6)2018 product type lead.Device code (b)(4) has been replaced with (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information from a health care provider (hcp) via a manufacturer representative (rep) clarified the lead/extension connector was suspected to have migrated.No further complications were reported/anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7454717
MDR Text Key106304704
Report Number3004209178-2018-08688
Device Sequence Number1
Product Code MRU
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2018
Device Model Number37603
Device Catalogue Number37603
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/20/2018
Initial Date FDA Received04/24/2018
Supplement Dates Manufacturer Received05/09/2018
Supplement Dates FDA Received05/11/2018
Date Device Manufactured12/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-