• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. INQUIRY AFOCUSII¿ (4 F DOUBLE LOOP) CATHETER 20 ELECTRODE, 7F; CATHETER, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. INQUIRY AFOCUSII¿ (4 F DOUBLE LOOP) CATHETER 20 ELECTRODE, 7F; CATHETER, ELECTRODE RECORDING Back to Search Results
Model Number IBI-87008
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/30/2018
Event Type  malfunction  
Manufacturer Narrative
One inquiry afocusii diagnostic catheter was received for evaluation.Visual inspection revealed a kink in the shaft noted 1.25¿ proximal to the sprial loop/shaft transition.Bends throughout the catheter shaft were noted proximal to the spiral loop/shaft transition.The catheter shaft had been twisted, had torn between 2.15¿ and 2.35¿ proximal to the spiral loop/shaft transition, exposing the internal spring body tube.The cause of the fractured and splitting the blue shaft was due to the catheter being over torqued during use.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
 
Event Description
During procedure, the surface material of the catheter was torn.The catheter was replaced and the procedure was completed with no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY AFOCUSII¿ (4 F DOUBLE LOOP) CATHETER 20 ELECTRODE, 7F
Type of Device
CATHETER, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7457422
MDR Text Key107085805
Report Number2030404-2018-00031
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K042775
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2020
Device Model NumberIBI-87008
Device Lot Number6102105
Other Device ID Number05414734309929
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/20/2018
Initial Date FDA Received04/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-