Product analysis: analysis found that the lower case and output connector assembly were broken on the external pulse generator (epg).Also, the display wires were pinched and the insulation was compromised.One of the capacitors on the main printed circuit board (pcb) was damaged.Additionally, the keypad was scratched.All found defective parts were replaced and all other identified issues were resolved.The device was re-calibrated and functionally tested.The device passed all final testing.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.If information is provided in the future, a supplemental report will be issued.
|