PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problem Output below Specifications (3004)
Patient Problem No Patient Involvement (2645)
Event Date 04/11/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) intermittently had a low atrial current at high settings.The issue did not occur at lower settings.The epg has been returned for service.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: analysis could not confirm the reported issue; output testing at multiple settings noted no anomalies.The external pulse generator (epg) failed incoming testing for atrial pace pulse noise.It was found that the upper and lower cases, display flex, display frame, keypad flex, main printed circuit board (pcb) and six main pcb screws were all contaminated.It was also noted that the display wires were pinched, but the insulation was not compromised.All found defective parts were replaced and all other identified issues were resolved.The epg passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7461978
MDR Text Key106611824
Report Number3004593495-2018-00403
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169725126
UDI-Public00643169725126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2018
Initial Date FDA Received04/25/2018
Supplement Dates Manufacturer Received05/08/2018
Supplement Dates FDA Received05/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-