Model Number S7 |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/04/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.Testing revealed that the reported issue could not be replicated.The system then passed the system checkout and was found to be fully functional.
|
|
Event Description
|
Medtronic received information that, while outside of a procedure, the navigation system became unresponsive while loading a cd.Restarting the navigation system restored functionality.There was no patient present when this issue was identified.No additional information was provided.
|
|
Manufacturer Narrative
|
A software investigation analysis was initiated to determine the probable cause of the issue through log analysis.Analysis found that the behavior exhibited was the intended function of the navigation system.As the disc could not be read by the navigation system and the system displayed an "unresponsive state".It was noted that the disc contained bad sectors, resulting in failed attempts to read.Analysis found that the software functioned as designed.
|
|