It was reported that the patient was undergoing generator replacement surgery when the surgeon identified abraded openings in the lead tubing (reported in mfr.Report #1644487-2018-00663), so a lead revision was then performed.It was then identified that there was no strain relief, and the electrodes were detached from the nerve.The surgeon and ear, nose, and throat specialist stated that the nerve was bubbled and severely scarred from the electrodes rubbing on it.The impedance was within normal limits prior to the surgery, which indicated that the electrodes were in contact with patient tissue while implanted.It was believed that the electrodes rubbed against the nerve after becoming detached, which caused the damage to the nerve.The lead was received by the manufacturer, and analysis was performed (reported in mfr.Report #1644487-2018-00663).However, the electrode portion of the lead was not returned, so no analysis could be performed on the portion of the lead relevant to this report.No further relevant information has been received to date.
|