• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problem Occlusion Within Device (1423)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/19/2018
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial# (b)(4), implanted: (b)(6) 2018, product type: catheter.Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), (b)(6), (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a health care provider (hcp) via a device manufacturer representative regarding a patient receiving bupivacaine (10 mg/ml at an unknown dose) and morphine ("2000 mcl/ml at 500 mcl/ml") via an implantable infusion pump.The indication for use was noted to be non-malignant pain.It was reported that the hcp believed that the catheter was "clogged" near the anchor.The majority of the catheter was removed and replaced with a new 8784 segment.The issue was diagnosed via dye study, which occurred "last week sometime." after replacement they were able to aspirate easily and a myelogram looked normal.No patient symptoms were reported.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7470129
MDR Text Key106795050
Report Number3004209178-2018-09529
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169530126
UDI-Public00643169530126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2019
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/26/2018
Initial Date FDA Received04/27/2018
Date Device Manufactured01/31/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age35 YR
-
-