PHILIPS HEALTHCARE ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number 722010
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
When the investigation has been completed philips will inform the fda.
 
Event Description
It was reported to philips that during a procedure on (b)(6) 2018 a patient went into cardiac arrest.This cardiac arrest was prior to an error message that appeared: ¿table movement only partially available".So this malfunction did not cause the cardiac arrest.The table did not lock and the customer performed a cardiopulmonary resuscitation.The customer finished the procedure successfully in another room on the same day.A new power supply was installed and the system is back in clinical use.
 
Manufacturer Narrative
Philips investigated this complaint.Philips checked the system and confirmed that the power supply of the table failed.The power supply was replaced and the system was put back in working order.Analysis of replacement rate for power supply did not show any negative trend.No further action will be taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS HEALTHCARE
veenpluis 4-6
p.o. box 10.000
best 5680 DA
NL  5680 DA
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
dusty leppert
veenpluis 4-6
p.o. box 10.000
best 5680 -DA
NL   5680 DA
MDR Report Key7471900
MDR Text Key106979391
Report Number3003768277-2018-00031
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number722010
Device Catalogue Number722010
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/10/2018
Initial Date FDA Received04/30/2018
Supplement Dates Manufacturer Received04/10/2018
Supplement Dates FDA Received07/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age71 YR
-
-