Model Number 16402
Device Problems Loose or Intermittent Connection (1371); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/05/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The field service representative (fsr) verified the reported complaint.He replaced the central processing unit (cpu) and the pump driver circuit board.The unit operated to the manufacturer's specifications.The suspect parts were returned to the manufacturer for evaluation.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, there was an "ovrspd1" error message displayed.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) placed the central processing unit (cpu) board on the 8k pump platter.The pump platter was powered on and rotations started.No overspeed error messages occurred and the pump rotated as expected.The pst tapped on the cpu board near u1 and u2 which caused an overspeed error to occur.After examining u1 and u2 it was found that u1 was not fully seated.U1 was pressed fully back into its socket and no further errors occurred.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 UNIVERSAL ROLLER PUMP
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
eileen dorsey
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key7472010
MDR Text Key106978938
Report Number1828100-2018-00216
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/06/2018
Initial Date FDA Received04/30/2018
Supplement Dates Manufacturer Received05/21/2018
Supplement Dates FDA Received05/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/15/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-