TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
|
Back to Search Results |
|
Model Number 16402 |
Device Problems
Loose or Intermittent Connection (1371); Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The field service representative (fsr) verified the reported complaint.He replaced the central processing unit (cpu) and the pump driver circuit board.The unit operated to the manufacturer's specifications.The suspect parts were returned to the manufacturer for evaluation.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, there was an "ovrspd1" error message displayed.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) placed the central processing unit (cpu) board on the 8k pump platter.The pump platter was powered on and rotations started.No overspeed error messages occurred and the pump rotated as expected.The pst tapped on the cpu board near u1 and u2 which caused an overspeed error to occur.After examining u1 and u2 it was found that u1 was not fully seated.U1 was pressed fully back into its socket and no further errors occurred.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|