(b)(4).Device evaluated by mfr: returned product consisted of a spiroflex thrombectomy system.The pump, effluent/supply line, shaft, tip, piston, and spike line were microscopically, tactile and visually inspected.There was a small amount of fluid in the boot when received.There were numerous kinks throughout the shaft of the device and in the spike line.Functional testing was performed by placing the device in the angiojet console.The pump body was loose in the capture block.Functional testing consisted of running the complaint device through the full priming cycle and 180 seconds in thrombectomy mode.The device ran within normal range without any issues or alarms/errors; however, there was a leak in the shaft.The shaft was inspected and it was revealed that 39cm distal of the strain relief, there was a hole in the pebax shaft.The piston head was manually rotated and ran through priming with issues or changes to the priming outcome.The od (outer diameter) of the piston rod (hex a) was measured and the dimension of the hex e was measured.Inspection of the device presented no other damage or irregularities.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|