• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Device Problems Difficult or Delayed Positioning (1157); Difficult to Insert (1316); Premature Activation (1484)
Patient Problem No Code Available (3191)
Event Date 03/25/2018
Event Type  Injury  
Manufacturer Narrative
If explanted, give date: not applicable as the lens remains implanted and therefore not explanted.(b)(6).Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site as it remains implanted; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that during the implantation of the intraocular lens (model pcb00) the cartridge tip was not fitting into the incision.Reportedly, the incision was done with a 2.4 mm knife.The incision had to be enlarged.The screw of the injector was not working unless the plunger was pushed.This caused the lens to come suddenly out of the delivery system.The lens remains implanted and the patient outcome was reported as good.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Type of Device
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
Manufacturer (Section G)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
pam mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key7478268
MDR Text Key107065459
Report Number2648035-2018-00639
Device Sequence Number1
Product Code HQL
UDI-Device Identifier05050474558342
UDI-Public(01)05050474558342(17)200825
Combination Product (y/n)N
PMA/PMN Number
P980040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 05/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date08/25/2020
Device Model NumberPCB00
Device Catalogue NumberPCB0000240
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/03/2018
Initial Date FDA Received05/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/25/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-