This report has been identified as b.Braun medical inc.Internal report number (b)(4).One (1) used transfer set, with packaging, was returned for evaluation.Visual examination of the transfer set noted no visual defects.The set was vacuum leak tested per specification with passing results.The set was leakage tested per specification with passing results.Additionally, lines 9 and 13 were each pressurized at 10 psi, and the valves were opened and closed.No leakage was noted.The iog review shows at around 10:51am a situation where the solution on valve 9 (mag sulfate) seems to have a lot of bubbles for a particular order.A manifold flush was done and then priming of several ingredients (station 6, station 7 three times, station 8, station 9 five times).This is believed to be when the odd behavior, was noticed which was described as seeping up valve 13.There is no attempt to pump valve 13 after this.There's no evidence of hanging the wrong thing on station 9 or station 13, and no occlusions or excessive bubbles other than during that one order, and the solutions do not seem to run out faster than expected.There is no clear reason why this event occurred.Based on the review of the software logs and a thorough investigation of the transfer set, this particular defect cannot be confirmed.If additional pertinent information becomes available a follow-up report will be filed.
|