• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750833
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A nurse manager reported that the foot pedal did not respond and the system was not able to perform phacoemulsification.There are no other details at this time.
 
Manufacturer Narrative
The company service representative examined the system, found the system, and associated footswitch to function as intended.The company service representative determined the issue to be due to nonconforming handpieces, unrelated to the system.The system was then tested and met all product specifications.The system was manufactured on april 29, 2008.Based on qa assessment, the product met specifications at the time of release.The footswitch serial number (s/n) was not provided and could not be determined based on the information provided.Therefore, manufacturing information could not be obtained.The system was found to meet specifications; therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key7481039
MDR Text Key107321735
Report Number2028159-2018-00907
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750833
Other Device ID Number380657508334
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/16/2018
Initial Date FDA Received05/02/2018
Supplement Dates Manufacturer Received08/20/2018
Supplement Dates FDA Received09/04/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
INFINITI FOOTSWITCH
Patient Outcome(s) Other;
-
-