ARTHROCARE CORP. EVAC 70 XTRA HP COBLATOR II; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Catalog Number EIC5874-01 |
Device Problems
Partial Blockage (1065); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/08/2018 |
Event Type
Injury
|
Event Description
|
It was reported that the probe could not be effectively cut and ablated, and the probe was severely blocked, so it was impossible to continue the operation.When the incident occurred, a simple check of the equipment had been performed and the saline control valve, pedals and connection lines were checked and no abnormalities were found.There was a delay greater than 30 minutes.
|
|
Manufacturer Narrative
|
The returned device, used in treatment, was returned as an mdr for evaluation.There was no relationship found between the device and the reported incident.Visual inspection under magnification of the wand shows minimal electrode wear and scratch/scuff marks on the return electrode and spacer.The electrodes are strongly contaminated with unknown substances.There are no manufacturing abnormalities visually observed with the returned wand.During functional evaluation in the lab the returned wand was activated in saline solution using a known good coblator ii controller.The eic device was activated in saline solution on ablate- and coag mode and creates plasma on the electrodes as intended.The suction- and the saline line were tested and performed as intended.The complaint was not verified as the returned device performed as intended.The root cause could not be determined with confidence as several factors could contribute to the reported event.Factors unrelated to the design and manufacture of the device which could have contributed to the complaint event; e.G.(1) ensure suction is properly connected, if not, stop and connect suction (2) evac wands are designed for ablation and/or coagulation only.Therefore based on insufficient information, a clinical assessment cannot be performed at this time.
|
|
Event Description
|
It was reported that the probe could not be effectively cut and ablated, and the probe was severely blocked, so it was impossible to continue the operation.When the incident occurred, a simple check of the equipment had been performed and the saline control valve, pedals and connection lines were checked and no abnormalities were found.The failure cause a patient thermal damage.
|
|
Search Alerts/Recalls
|
|
|