Model Number S7 |
Device Problem
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/10/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative was unable to replicate the reported issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced.No parts have been received by the manufacturer for evaluation.
|
|
Event Description
|
Medtronic received information that while outside of procedure, when the download button was clicked on a patient exam that the site was trying to load from a cd, the software unexpectedly exited to a blue screen.Pressing ctrl+alt+backspace to exit and relaunching the software resolved the issue and the site was able to download the exam successfully.There was no patient present at the time of the issue.
|
|
Manufacturer Narrative
|
The logs for the navigation system were reviewed by medtronic personnel.However, the logs provided no additional insight into the probable cause of the anomaly.
|
|
Manufacturer Narrative
|
A software investigation analysis was initiated to determine the probable cause of the issue through known anomaly determination.Analysis found that the behavior was consistent with a known software anomaly and that the issue was related to the software.If information is provided in the future, a supplemental report will be issued.
|
|