TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
|
Back to Search Results |
|
Model Number 16402 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed the black wire to be disconnected from the pump assembly when the housing was removed.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The product surveillance technician (pst) reported that during routine testing of the device at the service center, there was a loose wire on the pump cable assembly.There was no patient involvement.
|
|