Catalog Number 1012454-12
Device Problems Difficult to Remove (1528); Folded (2630)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The product was not returned to abbott vascular for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents reported from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Event Description
It was reported that the procedure was performed to treat a heavily calcified lesion in the right coronary artery (rca).The 4.5 x 12 mm nc trek balloon catheter winged out after inflation to 12 atmospheres and could not be pulled back into the guiding catheter.Both devices had to be removed as a single unit.There was no adverse patient effect.Although, this caused an inconvenience to the physician, there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7485433
MDR Text Key107451406
Report Number2024168-2018-03291
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648152108
UDI-Public08717648152108
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 05/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Catalogue Number1012454-12
Device Lot Number80206G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/23/2018
Initial Date FDA Received05/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial