• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC APPLICATION INSTRUMENT FOR STERNAL ZIPFIX; CERCLAGE FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC APPLICATION INSTRUMENT FOR STERNAL ZIPFIX; CERCLAGE FIXATION Back to Search Results
Model Number 03.501.080
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/11/2018
Event Type  malfunction  
Manufacturer Narrative
Device used for treatment, not diagnosis.Patient information not available for reporting.Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device history records review not available to report at this time.If additional is obtained that was not available for the initial med-watch, the investigation will be updated as applicable, and a follow-up med-watch will be filed as appropriate.
 
Event Description
It was reported that during a sternal debridement/reconstruction surgery on (b)(6) 2018, it was discovered that they could not fully tighten the zipfix implants because the internal mechanism which prevents the application instrument from over-tightening the zipfix implants was colliding with the cutting lever and pushing it forward.A new application instrument was used to complete the surgery.There was no harm to the patient and no surgery delay.Concomitant devices reported: zipfix implants (part unknown, lot unknown, qty unk).This complaint involves one (1) part.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A device history record review (dhr) was performed for the subject device: part: 03.501.080 , lot: 8726864 , manufacturing site: haegendorf , release to warehouse date: 29.Nov.2013.No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.An investigation was performed for the subject device: dhr review review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Service & repair evaluation the customer reported the internal mechanism was colliding with the cutting lever and pushing it forward.The repair technician reported the handle was loose, and the trigger was sticking.Lube/oil/clean is the reason for repair.The cause of the issue is unknown.No parts were replaced.The item was repaired per the inspection sheet, passed synthes final inspection and returned to the customer on 26-apr-2018.The evaluation was confirmed.The device was deemed serviceable and returned to the customer.No design or manufacturing issues were identified; therefore, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPLICATION INSTRUMENT FOR STERNAL ZIPFIX
Type of Device
CERCLAGE FIXATION
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7488563
MDR Text Key107921048
Report Number2939274-2018-52034
Device Sequence Number1
Product Code JDQ
UDI-Device Identifier10887587010847
UDI-Public(01)10887587010847
Combination Product (y/n)N
PMA/PMN Number
K110789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03.501.080
Device Catalogue Number03.501.080
Device Lot Number8726864
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2018
Initial Date Manufacturer Received 04/12/2018
Initial Date FDA Received05/04/2018
Supplement Dates Manufacturer Received06/05/2018
Supplement Dates FDA Received06/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-