Model Number 917000000
Device Problem Component Falling (1105)
Patient Problem No Patient Involvement (2645)
Event Date 04/11/2018
Event Type  malfunction  
Manufacturer Narrative
Maintenance review, investigation: the service technician visually inspected the machine at the customer's site and confirmed the reported condition.The technician found that the iv pole lever was slightly bent from inside the device.He straightened the iv pole lever and made slight adjustment to the iv pole set screw button present on the side panel of the machine and verified that iv pole is adjusted as per the manufacturer's specifications.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no problems identified related to the reported condition.Root cause: the root cause was the iv pole button set screw was out of alignment and the bent iv pole lever.Correction: a trima field action has been initiated to notify all trima users to use precaution while transporting the device and a caution statement was included in the operator's manual.Corrective action: an internal capa has been initiated to evaluate reports of the iv pole dropping down suddenly.
 
Event Description
The customer reported that the iv pole on the trima would not stay up.Per customer, no injuries occurred to the patient (donor) and operator was reported for this specific incident and therefore no patient (donor) or operator information is reasonably known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
steve kern
10810 w. collins ave.
lakewood, CO 80215
3032392246
MDR Report Key7488759
MDR Text Key107593669
Report Number1722028-2018-00119
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK010046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Notification
Type of Report Initial
Report Date 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number917000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 04/11/2018
Initial Date FDA Received05/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/27/2003
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-