Model Number 91496
Device Problems Device Inoperable (1663); Loss of Data (2903)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/16/2018
Event Type  malfunction  
Manufacturer Narrative
The uvsl command module was received at spacelabs¿ equipment service center for repair where the reported problem was verified.The printed circuit board assembly was replaced.The repaired unit passed all functional tests and was returned to the customer.This report is complete and this particular issue is considered closed.
 
Event Description
Spacelabs received a report on (b)(6) 2018, that while on a patient, a uvsl command module stopped functioning with loss of electrocardiogram and pulse oximetry data.No injury was reported with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACELABS ULTRAVIEW SL COMMAND MODULE
Type of Device
ULTRAVIEW SL MULTIPARAMETER MODULE
Manufacturer (Section D)
SPACELABS HEALTHCARE INC.
35301 se center st.
snoqualmie WA 98065
Manufacturer Contact
michael frey
35301 se center st.
snoqualmie, WA 98065
425 363-56
MDR Report Key7490423
MDR Text Key107733706
Report Number3010157426-2018-00069
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103142
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number91496
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/16/2018
Initial Date FDA Received05/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse