• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012452-08
Device Problems Deflation Problem (1149); Folded (2630)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/22/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported complaints.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that the procedure was to treat an 80% stenosed lesion in the circumflex.A 3.75x8mm nc trek balloon dilatation catheter (bdc) was being used to post-dilate a deployed non-abbott stent.The contrast mix was 50:50 and the balloon was inflated 2 times at 14 atmospheres.Difficulty deflating the balloon was noted and the balloon prolapsed on the proximal end when it was deflated.The bdc was removed with the balloon deflated.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7492041
MDR Text Key107588831
Report Number2024168-2018-03355
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648152016
UDI-Public08717648152016
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 05/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2020
Device Catalogue Number1012452-08
Device Lot Number70906G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/24/2018
Initial Date FDA Received05/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient Weight100
-
-