MAUDE Adverse Event Report: THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problems Occlusion Within Device (1423); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/20/2016
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction clot observed.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot e355 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot e355 for the reported issue shows no trends.Trends were reviewed for complaint categories, clot observed and alarm #17: return pressure.No trends were detected for each complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.(b)(4).
 
Event Description
The customer called to report a blood clot was observed during the treatment procedure.The customer stated they received an alarm #17: return pressure alarm and checked the patient's vascular access.The customer stated when flushing the patient's access a blood clot was pulled out of the return line.The product support specialist recommended the customer abort the treatment and not return any blood to the patient.The customer stated they would troubleshoot the patient access before deciding to continue with the procedure.The customer did not return product for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS, INC.
bedminster NJ 07921
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key7494653
MDR Text Key108161927
Report Number2523595-2018-00090
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)E355(17)180901
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date09/01/2018
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXUSA
Device Lot NumberE355
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/20/2016
Initial Date FDA Received05/08/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-